COOLING INCUBATOR:
Osworld Cooling Incubators are designed to serve a wide array of requirements, more specifically to conduct life cycle testing, shelf life studies, general incubation and refrigerated storage. These Cooling Incubators guarantee extremely accurate temperature regulation and uniform temperature distribution
- Temp. Control: Microprocessor-based PID Control with Auto-tune, CE Marked
- Temp. Sensor: PT-100 RTD, Class “A”
- Temp. Display: Digital LED, 3½ digit
- Heating: U-shaped Nichrome wire in SS Sheathing
- Cooling: CFC Free compressor utilizing R134-a eco friendly refrigerant, with condenser, motor and relay. Complete Unit – Copeland Make.
- Air Circulation: Flange motor with Impeller/ Blower
- Chamber Illumination: Fluorescent light with door switch
- Feet: Castor Wheels
- Construction: Double-walled with Poly Urethene (PUF) insulation provided with outer stainless steel metal door and inner glass viewing door
- Trays: SS Wire Mesh, heavy duty
Installation Mandatory Requirement
- Room temperature around machine should preferably be at 25°C with air conditioning or a well ventilated room with an exhaust fan. However, the surrounding temperature should not exceed 30°C.
2. Stabilized Input Voltage of 230V AC, 20 Amps. Use of a Servo Controlled Stabilizer is recommended.
i)Calibration
- Temperature sensor probe calibration with traceability to ERTL
- Temperature controller calibration with traceability to ERTL
- ii) Validation
- IQ, OQ and PQ documentation and protocols shall be provided with equipment.
- Documentation prepared / developed for Osworld by Premier Validation Ltd. – Ireland, one of Europe’s leading validation consultants
- Performance Qualification with Temperature Mapping on site. The Performance Validation Test consists of one cycle at any one temperature point for an 8-hour period. Data is collected in printed format from up to six different points within the chamber. Temperature uniformity is considered acceptable if the deviation is less than ±1°C of set temperature.
OPTIONAL EXTRA/ACCESSORIES:
- Oslog Data Storage Device and Oslog 21 CFR Part 11 Compliant Software – Both have been developed exclusively for Osworld by M/s Godrej & Boyce Mfg. Co. Ltd.
- Additional high-temperature safety system to cut off all heaters with audio-visual alarm
- PLC-based system: PLC with HMI 3.7” touch screen grey monitor with 21 CFR Part 11 features (OSLOG) software.
- Door access security system
- Voltage Stabilizer for above unit
- GSM System: Alarms sent to predefined mobile numbers
- HMI: 5.7″ touch screen grey in lieu of 3.7″
- Controller with printer interface to connect to EPSON Dot Matrix line printer. Date, time and temperature can be printed whereas print interval can be set between 1 to 99 minutes
- Data Logger: 4 point temp. Data logger with sensors placed at specific points in chamber with printer interface and memory